Agilent / HP / Keysight 8010A
8010A PULSE GENERATOR
Description
The Agilent/HP 8010A is a dual channel pulse generator which has a repetition rate of 1 Hz to 10 MHz, which is extendable to 20 MHz in double pulse mode.
Individual switching of channels A and B permits the application of an external trigger to one channel while the other remains internally triggered.
Complete regulation of the pulse delay, width, rise and fall times, amplitude and dc offset can be achieved by front panel controls.
All the front panel controls of HP 8010 are calibrated with the exception of the dc offset.
In the parallel operation of the normal delay mode, both pulses are referenced to the trigger output, therefore, optional delay of either pulse can be obtained while the other remains fixed.
Hence, the output of channel A or B can be delayed from 50 ns to 1 s with respect to the trigger output.
In the series mode, combining A and B makes channel A the reference for channel B. Channel amplitudes are adjustable independently from 5 volts to 0.02 volts with consideration for the output impedance of 50 ohms, a seven step attenuator of HP 8010A reduces the output in a 5, 2.5 and 1 sequence.
A maximum of 10 volts output amplitude is readily available from channel B, by operation of a rear panel switch and the output combining switch.
The pulse parameters are consequently controlled through channel B. The 8010A pulse generator can be operated in a square wave mode.
Symmetrical, dc coupled, square waves from 1 Hz to 10 MHz with variable rise and fall times are available at both outputs.
Key Points
Specifications
Power Requirements: 115 V or 230 V + 10%, -15%, 48 Hz to 440 Hz.
Weight: Net 23.45 lb (10.7 kg).
Dimensions: 16-3/4” wide, 7-1/4” high, 18-3/8” deep overall (425 x 184 x 466 mm).
PULSE CHARACTERISTICS: (50 ohm load impedance)
Rise and Fall times
Sep. Outputs: < 10 ns to 1 second in eight ranges; ranges are common for rise and fall times. Independent verniers provide separate control of rise and fall times within each range up to a maximum ratio of 1 : 10.
Combined Outputs: < 12 ns to 1 second
10 V Output: < 20 ns to 1 second
Accuracy: ±10% of setting, ±2% of full scale ±4 ns.
Linerarity: For transition time > 30 ns maximum amplitude deviation from a straight line between the 10% and 90% points is less than 4% of pulse amplitude.
Overshoot and Ringing: < 5% of pulse amplitude.
Pulse Width: < 20 ns to 1 second in eight ranges. Vernier provides continuous adjustment between ranges.
Accuracy: ± 10% of setting, ± 2% of full scale ± 4 ns.
Maximum Duty Cycle: > 80% for repetition rates from 1 Hz to 1 MHz; > 50% from 1 MHz to 10 MHz.
Width Jitter: < 0.1% on any width setting.
Maximum Output: 5 V separate or combined outputs, 10 V channel B (channel A no output).
Attenuator: Seven-step attenuator reduces output to 0.05V in 5, 2.5, 1 sequence. Vernier provides continuous adjustment between steps and reduces minimum output to < 0.02 V.
Accuracy: ± 10% of setting ± 2% of full scale.
Pulse Outputs:
Sep. Outputs: Two outputs, each positive or negative selectable.
Comb. Outputs: Outputs of channel A and B internally added (no loss of amplitude).
Outputs short-circuit proof.
Source Impedance: 50 ohm ± 10% shunted by typically 20 pF.
DC Offset: ±2 V across 50 ohm load. Independent of attenuator and vernier setting; can be switched off.
Pulse Delay (Parallel): < 80 ns to 1 second delay with respect to trigger output. Eight ranges. Vernier provides continuous adjustment between ranges.
Serial: Max. delay for channel B can be doubled.
Accuracy: ±10 % of setting, ± 2% of full scale ± 4 ns.
Delay Jitter: < 0.1% on any delay setting.
REPETITION RATE AND TRIGGER
Free Running: 1 Hz to 10 MHz in seven ranges. Vernier provides continuous adjustment between ranges.
Accuracy: ±10% of setting, ±2% of full scale.
Period Jitter: < 0.1%
Square wave: 1 Hz to 10 MHz output symmetrical to ground, rise and fall times variable, 50% duty cycle.
Double pulse: Minimum pulse distance of < 50 ns allows maximum rep. rate up to 20 MHz.
External Triggering: Repetition Rate - 0 to 10 MHz. Can be triggered with sines waves or pulses of either polarity. (For Square Wave Output frequency divided by a factor of 2).
Trigger Input: Sine waves 1 V p-p. Pulses 0.5 V peak at least 20 ns wide, maximum 10 V.
Delay: Approximately 30 ns between trigger input and trigger output.
Input impedance: 1.0 k ohm
Manual: Push button for single pulse.
Separate Triggering of channel A or B: Spikes + 2 V amplitude < 50 ns width. Input impedance 50 ohm (inputs on rear panel).
Trigger Output: Amplitude < + 2 V across 50 ohm from 50 ohm, Width: 15 ns ± 10 ns.
GATING: Synchronous Gating - Gating signal turns generator “on”, Asynchronous Gating - Gating signal turns the output pulse “on”. Trigger output always available.
Gate inputs: At least -8 V enabling to –10 V.
Operating Temperature: 0 deg C to + 55 deg C.