Agilent / HP / Keysight 6920B
6920B METER CALIBRATOR
Description
This unit is a feedback regulated power source suited for the calibration of meters.
It has ten continuously variable output ranges ranging from 1 Volt to 1000 volts and from 100uA to 5A full output.
All ranges provide DC or AC at power line Frequency but the 100uA range is not calibrated in AC.
This compact, solid-state calibrator was designed for your convenience, and combines in one instrument all the outputs needed to test the more commonly used meters.
Capable of constant-voltage and constant-current output, its absolute accuracy (±0.2% DC, ±0.4% AC) makes it suitable for laboratory or production testing.
Key Points
AC or DC output
Voltage: .01 to 1000 Volts in 4 ranges plus vernier
Current: 1 uA to 5 Amps in 6 ranges plus vernier
0.5% or better accuracy
Momentary and toggle on output switch
Specifications
DC Voltmeters up to 1000 V
Average reading AC Voltmeters up to 1000 V
DC Ammeters up to 5 amps
Average reading AC ammeters up to 5 amps