Agilent / HP / Keysight 6632A
6632A GPIB dc power supply, 0-20 Vdc, 0-5 A
Description
The Agilent / HP 6632A is 100 W, single output, series pass regulated power supply optimized for system applications.
The 6632A power supply provides a high performance output with a full set of programmable features.
The Agilent 6632A is a one-box solution for system power supply requirements. Both the voltage and current output can be programmed for either CV or CC operation.
A built-in DVM and precision current shunt measure (readback) the actual power supply output.
The internal HP-IB interface allows for complete programmability, including status reporting and interrupt generation with user designated fault conditions.
In addition, programming commands are easy-to-use, and self-documenting.
The Agilent / HP 6632A sink as well as source current. This provides very fast down programming times and speeds throughput in production test applications. The negative current setting tracks the user-programmed positive current (CC) setting.
Remote sense terminals allow for precise voltage regulation at the load. This feature will compensate for load lead IR drops of up to 2 volts per lead.
The Agilent / HP 6632A has Overvoltage and Overcuurent protection. Both features are programmable via the front panel keypad and HP-IB. The Overvoltage protection includes an SCR crowbar.
The Agilent 6632A has built-in selftest capability. This feature verifies the function of all HP-IB circuitry, D/A's and A/D's at power on and upon HP-IB command.
Software calibration eliminates shock sensitive potentiometers normally used for calibration of power supplies and programmers. This feature enables calibration of the Agilent 6632A power supply via the HP-IB while mounted in the rack.
Option 020 adds a front panel output in parallel with the rear panel output terminal strip. These front panel binding posts provide flexibility in accessing the power supply output, and are very useful in bench applications and system development.
Key Points
100 Watt Linear Output
Current Sinking Capability
Built-in HP-IB interface
Overvoltage & Overcurrent Protection
Readback Measured Voltage and Current Output via HP-IB
Linear output regulation
Low ripple and noise
"One-Box" solution: includes V and I readback
Fast up and down programming
Specifications
Output ratings: Output voltage: 0 to 20V, Output Current: 0 to 5A
Programming Accuracy: Voltage: 0.05% + 10mV, Current: 0.15% +7mA
Ripple and noise from 20Hz to 20MHz: Voltage (rms/peak-peak): Normal mode: 0.3mV/10mV, Fast mode: 1mV/10mV, Current: RMS: 2mA
Readback Accuracy: Voltage: 0.07% + 15mV, +Current: 0.18% +9mA, -Current: 0.50% + 15mA
Load regulation: Voltage: 2mV, Current: 1mA
Line regulation: Voltage: 0.5mV, Current: 0.5mA