Agilent / HP / Keysight 4395A
4395A Network/Spectrum/Impedance Analyzer
Description
The 4395A provides excellent vector network, spectrum, and optional impedance measurements for lab and production applications. Gain, phase, group delay, distortion, spurious, CN, and noise measurements often required for evaluating components and circuits can be measured using one instrument.
When combined with a test set, the 4395A provides reflection measurements, such as return loss, SWR, and S parameters. As a vector network analyzer, the 4395A operates from 10 Hz to 500 MHz with 1 mHz resolution; and its integrated synthesized source provides -50 to + 15 dBm of output power with 0.1 dB resolution. The dynamic magnitude and phase accuracy are +/-0.05 dB and +/-0.3 deg. so that it can accurately measure gain and group delay flatness, which are becoming more important in modern electronics systems.
As a spectrum analyzer, the 4395A operates from 10 Hz to 500 MHz with resolution bandwidths (RBWs) spanning 1 Hz to 1 MHz in a 1-3-10 steps. A fully-synthesized local oscillator allows stable and accurate frequency analysis. Direct A/D conversion (no LOG amplifier is used) results in +/-0.8 dB level accuracy (@50 MHz, -20 dBm). Noise sidebands fall below -100 dBc/Hz @ 100 kHz offset from carriers, while sensitivity is -145 dBm/Hz at 10 MHz.
Key Points
Full-vector network and spectrum measurement and analysis
Wide dynamic range network measurement with fast sweep speeds
Extremely fast narrowband spectrum measurement
Impedance analysis option and test kit available
-145dBm/Hz sensitivity for spectrum analysis
Built-in IBASIC for easy test automation
Time-gated spectrum analysis option
Color TFT display and built-in disk drive/RAM disk
Specifications
Network analyzer specifications:
Frequency range: 10Hz to 500MHz
Frequency resolution: 1mHz
Output power range: –50 to +15 dBm
Dynamic range: 11 dB @ 10Hz IFBW
Dynamic accuracy: ±0.05 dB/0.3 deg.
Calibration: Full two-port
Spectrum analyzer specification:
Frequency range: 10Hz to 500MHz
Noise sidebands: < -104 dBc/Hz, typical at 10KHz offset
Resolution bandwidth: 1Hz to 1MHz in 1-3-10 steps
Dynamic range: > 100 dB third order free dynamic range
Level accuracy: ±0.8 dB @ 50MHz
Sensitivity: -145 dBm/Hz @ freq. = 10MHz
Impedance analyzer specification:
Frequency range: 100KHz to 500MHz
Meas. parameter: |Z|, ?z, R, X, |Y|,?y, G, B, Cs, Cp, Ls,Lp, Rp, Rs, X, D, Q,I?I, ?x ,?y
Z accuracy: ±3% (typical basic accuracy)
Source level: -56 dBm to +9 dBm (at DUT)
DC bias: ±40 V (20 mA (max)
Compensation: OPEN/SHORT/LOAD port extension
Output Impedance: 50 ohms
Weight: 21.5 kg (47 lb.)