Description
The 4287A RF LCR meter offers accurate, reliable and fast measurements from 1 MHz to 3 GHz to improve quality and throughput of electronic component testing in production lines.
The 4287A employs the direct-current voltage-measurement technique, as opposed to the reflection-measurement technique, which yields accurate measurements over a wide impedance range.
Key Points
1 MHz to 3 GHz, with 100 kHz steps
Wide impedance measurement range from 200 m ohm to 3 k ohm
1% basic accuracy
Superior measurement repeatability at low test signal level
High-speed measurements: 9 ms
Specifications
Frequency:
Range:1 MHz to 3 GHz
Resolution:100 kHz
Uncertainty:
± 10 ppm (23 °C ± 5 °C)
± 20 ppm (5 °C to 40 °C)
Oscillator level:
Resolution: 0.1 dB
Power (when 50 ohm LOAD is connected to the test port):
–40 dBm to 1 dBm (Frequency >= 1 GHz)
–40 dBm to 0 dBm (Frequency > 1 GHz )
Current (when SHORT is connected to the test port):
0.0894 mArms to 10 mArms (Frequency >= 1 GHz)
0.0894 mArms to 8.94 mArms (Frequency > 1 GHz )
Voltage (when OPEN is connected to the test port):
4.47 mVrms to 502 mVrms (Frequency >= 1 GHz)
4.47 mVrms to 447 mVrms (Frequency > 1 GHz )
Frequency <= 1 GHz:
± 2 dB (23 ± 5 ºC)
± 4 dB (5 ºC to 40 ºC)
Frequency > 1 GHz:
± 3 dB (23 ± 5 ºC)
± 5 dB (5 ºC to 40 ºC)
Output Impedance:
Output impedance: 50 ohm (nominal)
Environment conditions:
Operating condition:
Temperature:5 °C to 40 °C
Humidity (at wet bulb temperature = 29 °C, without condensation):15% to 90% RH
Warmup time :30 minutes
Non-operating storage condition:
Temperature :– 20 °C to + 60 °C
Humidity (at wet bulb temperature = 45 °C, without condensation):15% to 90% RH
Other specifications:
Power requirement:90 V to 132 V, or 198 V to 264 V (automatically switched), 47 Hz to 63 Hz, 350 VA max.
Weight:16 kg (SPC)