Agilent / HP / Keysight 400FL
400FL AC VOLTMETER
Description
This AC Voltmeter is primarily intended as a dB Meter, however it can indicate both V and dB.
The meter has a linear dB scale with a logarithmic AC scale below.
The voltage range is 100µV to 300V full scale.
Meter calibration responds to the absolute average value of the applied signal, calibrated in RMS Volts from 20Hz to 4MHz.
Key Points
Linear dB scale with a logarithmic AC scale below
Voltage range is 100µV to 300V full scale
Meter calibration responds to the absolute average value of the applied signal, calibrated in RMS Volts from 20Hz to 4MHz
Accuracy as a % of reading is ±4% (20 to 40Hz), ±2% (40 to 100Hz), ±1% (100Hz to 1MHz), ±2% (1 to 2MHz), ±4% (2 to 4MHz)
Input impedance is 10Mohms
AC amplifier output is 1V rms open circuit for full scale meter indication
Specifications
Voltage range: 100µV to 300V
Frequency range: 20Hz to 4MHz
<
p>Input impedance: <30pF to <15pF