Agilent / HP / Keysight 8565EC
8565EC Portable Spectrum Analyzer, 9 kHz to 50 GHz
Description
The Agilent 8565EC portable, color display millimeter wave spectrum analyzer offers the measurement capability and performance previously found in larger, more expensive benchtop analyzers. You can measure signals from 30 Hz to 50 GHz (preselected above 2.75 GHz) with a single coaxial connection. The 8565EC combines outstanding phase noise, sensitivity, 1 Hz resolution bandwidths, synthesized tuning and wide dynamic range in a package built to withstand environmental conditions. It offers a complete solution for adjacent-channel power (ACP) testing of burst carrier signals using digital modulation, and the ability to measure occupied bandwidth percentage.
Key Points
Outstanding performance for sensitivity, phase noise, the 2nd and 3rd order dynamic range, and the voltage standing wave ratio (VSWR)
Color Display
Continuous 30 Hz to 50 GHz sweep
Fast digital resolution bandwidths of 1, 3, 10, 30 and 100 Hz
Adjacent channel power, channel power, carrier power, occupied bandwidth percentage and time-gated measurements standard
Precision timebase and 1 Hz counter resolution
MIL-PRF-28800 Class 3 rugged
Measurement personalities for phase noise and spurious response measurements, more details
Easily transfer screen image or trace data to PC with E4444A BenchLink software
Specifications
Frequency Specifications:
Frequency range:
Internal mixing:30 Hz to 50 GHz
External mixing:18 GHz to 325 GHz
Sweep time:
Range
Span = 0 Hz: 50 µs to 6000 s
Span ±100 Hz:
RBW = 300 Hz: 50 ms to 2000 s
RBW = 100 Hz: 50 ms to 100 ks
Resolution bandwidth:
Range (–3 dB):1 Hz to 1 MHz in a 1, 3, 10 sequence and 2 MHz (3 MHz at –6 dB)
Video bandwidth range: 1 Hz to 3 MHz in a 1, 3, 10 sequence
Noise sidebands:
100 Hz :<= –88 dBc/Hz
1 kHz :<= –97 dBc/Hz
10 kHz :<= –113 dBc/Hz
30 kHz :<= –113 dBc/Hz
100 kHz :<= –117 dBc/Hz
Amplitude Specifications:
Range:Displayed average noise level (DANL) to +30 dBm
Maximum safe input level:
Average continuous power:+30 dBm (1 W, input attn = 10 dB)
Peak pulse power:+50 dBm (100 W, input attn = 30 dB)
Maximum DC input voltage:
DC coupled: ±0.2 Vdc
AC coupled: ±50 Vdc
Displayed average noise level:
30 Hz:<= –90 dBm
1 kHz:<= –105 dBm
10 kHz:<= –120 dBm
100 kHz:<= –120 dBm
1 MHz to 10 MHz:<= –140 dBm
10 MHz to 2.9 GHz:<= –145 dBm
2.9 GHz to 6.46 GHz:<= –147dBm
6.46 GHz to 13.2 GHz:<= –143 dBm
13.2 GHz to 22.0 GHz:<= –140 dBm
22.0 GHz to 26.8 GHz:<= –136 dBm
26.8 GHz to 31.15 GHz :<= –139 dBm
31.15 GHz to 40 GHz:<= –130 dBm
40 GHz to 50 GHz:<= –127 dBm
Dynamic range:
10 MHz to 2.9 GHz: > 144 dB
2.9 GHz to 6.46 GHz:> 148 dB
6.46 GHz to 13.2 GHz:> 142 dB
13.2 GHz to 22.0 GHz: > 137 dB
22.0 GHz to 26.8 GHz: > 136 dB
26.8 GHz to 31.15 GHz: > 139 dB
31.15 GHz to 40 GHz: > 130 dB
40 GHz to 50 GHz:> 127 dB
Environmental Specifications:
Dimensions (w/o handle, cover): 337 mm W x 187 mm H x 461 mm D
Calibration interval:2 years
Warm-up time :5 minutes in ambient conditions
Temperature: 0 °C to +55 °C (operating); –40 °C to +75 °C (not operating)
Humidity:95% @ 40 °C for 5 days
Maximum power dissipation:260 W
Weight (nominal):17.3 kg (38 lbs)
Power requirements:
115 VAC operation: 90 to 140 V rms, 3.2 A rms max, 47 to 440 Hz
230 VAC operation: 180 to 250 V rms, 1.8 A rms max, 47 to 66 Hz