Description
This Resistivity Cell is designed to operate specifically with the Agilent 4339A/B High Resistance Meter.
It is used to measure surface or volume resistance/resistivity of insulation materials.
Volume resistivity is measured across the material (MUT) by the unguarded and guarded electrodes.
Surface resistivity, on the other hand, is measured along the surface of the MUT (between the guarded and the unguarded electrodes).
Key Points
Three kinds of electrodes (diameter: 26mm/50mm/76mm), provided for the 16008B, can satisfy various insulation measurement standards such as ASTM D 257.
Triaxial input terminal configuration minimizes the influences due to external noise and as a result, high resistance up to 1.6E 16ohms can be measured accurately.
OPEN compensation function minimizes the influences due to leakage current.
Specifications
Maximum voltage: 1000V
Maximum current: 10mA
Operating temperature: -30 to +100 degrees C
DUT connection: 2 terminal
Cable length: 1.2m