Agilent / HP 10073B
Passive Probe
Description
The Agilent 10073B is a passive probe. It is a 500 MHz probe. It has a 10:1 attenuation ratio and high 2.2 M ohm input impedance. It is a general purpose high bandwidth passive probe for the Agilent 54615B and for the 54616B/C oscilloscopes. The probe identification feature will automatically configure the oscilloscope for the probe attenuation.
Key Points
Specifications
Bandwidth: 500 MHz
System risetime (calculated): <700 ps
Attenuation Ratio: 10:1
Input resistance: 2.2 MΩ (when terminated into 1 MOhm)
Input capacitance: ~12 pF
Maximum input (dc + peak ac): 500 V CAT I (mains isolated) 300 V CAT II (post receptacle mains)
Compensation range: 6 - 15 pF