Agilent / HP / Keysight N1911A
N1911A P-Series Single Channel Power Meter
Description
When this unit is used with the N192XA wideband power sensors, this P-Series single channel power meter has a measurement frequency range from 50MHz to 40GHz with an internal zero and calibration capability.
Key Points
30 MHz video bandwidth
Single shot real time capture at 100 Msample/s per second
Key Measurements: - peak, average, peak-to-average ratio, rise time, fall time and pulse width
22 Predefine formats: WiMAX, DME, HSDPA, etc.
One screen view for pulse measurement analysis: Auto Scale, Auto Gate, Rise/Fall Time, Duty Cycle, etc.
Internal Zeroing and Calibration while connecting to the DUT
LXI-C Compliance
Average triggering for average power measurement
Versatile connectivity with USB, LAN and GPIB standard
High Speed Complementary Cumulative Distribution Function (CCDF) statistical analysis
Compatible with the industry standard 8480 and E-series sensors which provide a wide dynamic range from -70 to +44 dBm, and a frequency coverage from 9 kHz to 110 GHz.
Compatibility with E932x power sensors
Difference and ratio math functions : (A-B, B-A, A/B and B/A)
2-year calibration cycle
Specifications
Number of channels:single channel
Measurements: Average, peak and peak-to-average ratio power measurements are provided with free-run or time-gated definitions.
Time parameter measurements of pulse rise time, fall time, pulse width, time-to-positive occurrence and time-to-negative occurrence are also provided.
Maximum sampling rate: 100 Msamples/sec, continuous sampling
Video bandwidth: = 30 MHz
Single-shot bandwidth: = 30 MHz
Rise time and fall time: = 13 ns (for frequencies = 500 MHz)
Minimum pulse width: 50 ns
Overshoot: = 5 %
Dynamic range: –35 dBm to +20 dBm (> 500 MHz); –30 dBm to +20 dBm (50 MHz to 500 MHz)
Maximum capture length: 1 second
Maximum pulse repetition rate: 10 MHz (based on 10 samples per period)
Timebase
Timebase range: 2 ns to 100 msec/div
Accuracy: ±10 ppm
Jitter: = 1 ns
1 mW Power Reference
Power output: 1.00 mW (0.0 dBm). Factory set to ± 0.4 % traceable to the National Physical Laboratory (NPL) UK
Accuracy (over 2 years): ±1.2 % (0 to 55 ºC); ±0.4 % (25 ± 10 ºC)
Frequency :50 MHz nominal
Connector type: Type N (f), 50 W
Net weight:= 3.5 kg (7.7 lb) approximate