Fluke 5500A-COIL
Coil 50 Turns
Description
This Coil is used as a tool for calibrating clamp type current meters that operate by two different principles ¾ as current transformers (AC only), and by the Hall Effect (Both AC and DC).
It is impractical to calibrate 1000A rated current clamp meters using a 1000A source.
However, by using the 50 turns of this unit in conjunction with a current calibrator you can effectively multiply the current of the current calibrator by a factor of 50 to support the calibration and verification of these clamp-on current meters.
Key Points
50 Turns
11A rms, continuous Maximum Current
3V rms Maximum Voltage
Specifications
Number of Turns: 50
Maximum Current: 11A rms, continuous
Maximum Voltage: 3V rms