Agilent / HP / Keysight MSO7034A
MSO7034A Mixed Signal Oscilloscope: 350 MHz, 4 analog plus 16 digital channels
Description
Agilent’s InfiniiVision DSO7034A, equipped with a 12.1” XGA display, comes in a quiet package that is just 6.5” deep, and weighs only 13 pounds. With the world’s fastest uncompromised update rate, up to 100,000 waveforms per second, see subtle signal detail and infrequent events other scopes miss. A wide range of application packages provide rapid measurement insight.
Key Points
Lab scope performance
350 MHz bandwidth
4 analog plus 16 digital channels
2 GSa/s sample rate
Standard 8 Mpts MegaZoom III deep memory
Engineered for the best signal visibility
12.1” XGA display with 256 levels of intensity shows subtle signal detail
Industry’s-fastest uncompromised update rate of up to 100,000 waveforms/sec displays infrequent events
Wide range of application packages
Serial decode including: I2C, SPI, CAN, LIN, RS-232/UART, and I2S
Core-assisted FPGA debug for Xilinx and Altera devices
Segmented memory
PC-based offline analysis and viewing of previously acquired data
Specifications
Acquisition: scope channels
Sample rate:2 GSa/sec each channel
Memory depth Standard: 2 channels/4 channels; 8 Mpts/4 Mpts
Vertical resolution: 8 bits
Peak detection:500-ps peak detect
Averaging: Selectable from 2, 4, 8, 16, 32, 64 … to 65536
High resolution mode: Average mode with avg = 1; 12 bits of resolution when =10 µs/div at 4 GSa/s or =20-µs/div at 2 GSa/s
Filter: Sinx/x interpolation (single shot BW = sample rate/4 or bandwidth of scope, whichever is less) with vectors on and in real-time mode
Acquisition: digital channels
Sample rate: 2 GSa/sec one pod**, 1 GSa/sec each pod
Maximum input frequency: 250 MHz
Memory depth Standard:One pod/both pods (with scope channels turned off) Standard 8 Mpts/4 Mpts
Vertical resolution: 1 bit
Glitch detection: 2 ns (min pulse width)
Vertical system: scope channels
Scope channels:Ch 1, 2, 3 and 4 simultaneous acquisition
Bandwidth (–3 dB):DC to 350 MHz
AC coupled:3.5 Hz to 350 MHz
Calculated rise time (=0.35/bandwidth):1 nsec
Single-shot bandwidth:350 MHz
Range:2 mV/div to 5 V/div (1 MO or 50 O)
Offset range: 2 mV/div to 5 V/div (1 MO or 50 O)
Dynamic range: ±8 div
Input impedance 1 MO ± 1% || 14 pF or 50 O ± 1.5%, selectable
Coupling :AC, DC
BW limit :25 MHz selectable
Channel-to-channel isolation: DC to max bandwidth >40 dB
Standard probes: 10073C shipped standard for each scope channel (1165A probes optional)
Probe ID: Auto probe sense and AutoProbe interface; Agilent- and Tektronix-compatible passive probe sense
ESD tolerance: ±2 kV
Noise, RMS, input shorted:0.50% FS or 300 µV, whichever is greater
Power requirements
Line voltage range: 100-120V, 50/60/400 Hz; 100-240V, 50/60 Hz auto ranging
Line frequency: 50/60 Hz, 100-240 VAC; 400 Hz, 100-120 VAC
Power usage: 120 W max