Agilent / HP 85033C
Calibration Kit 3.5mm
Description
The Agilent 85033C 3.5 mm calibration kit is designed to be used with network analyzer systems such as the Agilent 8752A or 8753A/B/C. The standard 85033C kit consists of opens, shorts, 50 ohm terminations and 3.5 mm to 7 mm adapters.
The network analyzer system usually provides prompts to help the user through a measurement calibration. Note that the network analyzer may be compatible with several different calibration kits. Make sure the 3.5 mm calibration kit is selected before starting a calibration with the 85033C.
The open standards in the Agilent 85033C Calibration kit consist of two parts. The open circuit outer conductor and the center conductor extender need to be used together. When using the open standard, center conductor extender must be selected carefully which will mate correctly with the test port connector.
Temperature of the calibration device is important because device dimensions and therefore electrical characteristics change with temperature. The temperature of the calibration devices and all connectors must be stable before use and within the operating tolerances.
Key Points
3.5mm Calibration Kit
Pha match 7mm to 3.5mm adpters
Specifications
Output Impedance: 50ohms
Type: Calibration Kit
Impedance: 50ohms
Connector Type to DUT: 3.5mm(m)
Grade: Precision
Net weight: 0.8 kg (1.8 lb)
Shipping weight: 1.4 kg (3.1 lb)
Operating temperature: +15 to +35 deg C
Storage temperature: -40 to +75 deg C